ISO 3567:2026
ISO 3567:2026 Vacuum gauges — Calibration by direct comparison with a reference gauge
CDN $192.00
This publication was last reviewed and confirmed in 2026.
Vacuum gauges — Calibration by direct comparison with a reference gauge
Description
This document specifies the physical, technical and metrological conditions to be fulfilled when calibrations of vacuum gauges are performed by direct comparison with a reference gauge at a calibration laboratory, whether accredited or not. From the conditions described, the design of an apparatus that can perform vacuum gauge calibrations in an adequate manner can be deduced.
This document is applicable for calibration of all types of vacuum gauges. Vacuum gauges can consist of several parts; typically, these are: gauge head, cable, and control unit. This entire set is considered as the unit to be calibrated. If only the gauge head (i.e. the part of the vacuum gauge directly exposed to the vacuum) is calibrated, all set-ups and conditions are recorded such that the user of the calibrated gauge head is able to perform the measurements in the same manner as during the calibration.
This document does not give guidance on how to treat special types of vacuum gauges, be they reference standards or units under calibration.
The applicable pressure range for calibrations treated in this document depends on the realized design of the calibration apparatus and on the type of reference gauge, and the range varies in its limits from 10−6 Pa to 133 kPa.
Edition
2
Published Date
2026-09-16
Status
PUBLISHED
Pages
15
Format 
Secure PDF
Secure – PDF details
- Save your file locally or view it via a web viewer
- Viewing permissions are restricted exclusively to the purchaser
- Device limits - 3
- Printing – Enabled only to print (1) copy
See more about our Environmental Commitment
Abstract
This document specifies the physical, technical and metrological conditions to be fulfilled when calibrations of vacuum gauges are performed by direct comparison with a reference gauge at a calibration laboratory, whether accredited or not. From the conditions described, the design of an apparatus that can perform vacuum gauge calibrations in an adequate manner can be deduced.
This document is applicable for calibration of all types of vacuum gauges. Vacuum gauges can consist of several parts; typically, these are: gauge head, cable, and control unit. This entire set is considered as the unit to be calibrated. If only the gauge head (i.e. the part of the vacuum gauge directly exposed to the vacuum) is calibrated, all set-ups and conditions are recorded such that the user of the calibrated gauge head is able to perform the measurements in the same manner as during the calibration.
This document does not give guidance on how to treat special types of vacuum gauges, be they reference standards or units under calibration.
The applicable pressure range for calibrations treated in this document depends on the realized design of the calibration apparatus and on the type of reference gauge, and the range varies in its limits from 10−6 Pa to 133 kPa.
Previous Editions
Can’t find what you are looking for?
Please contact us at:
Related Documents
-

ISO 19685:2017 Vacuum technology – Vacuum gauges – Specifications, calibration and measurement uncertainties for Pirani gauges
CDN $192.00 Add to cart -

ISO 20146:2019 Vacuum technology – Vacuum gauges – Specifications, calibration and measurement uncertainties for capacitance diaphragm gauges
CDN $192.00 Add to cart -